Single Run Analysis

Detailed procedure for running one sample on the Shimadzu GC-2025. For the abbreviated version, see Run One Sample. For automated sequences, see Batch_Processing.

Prerequisites: GC is started and showing “Ready” (see Startup), method file is loaded (see Method_Development).


Step 1: Sample Login

  1. Click Single Run icon on the Assistant Bar
  2. Click Sample Login icon

📋 From manual


Step 2: Enter Sample Information

FieldRequired?Details
Data FileMandatoryName for data file. Check Auto increment to avoid overwriting previous runs.
Vial #MandatorySample vial position on auto-injector. Use “0” for blank/cleanup runs.
Sample NameOptionalFor record-keeping
Sample IDOptionalFor record-keeping
ReportOptionalCheck to auto-print report after analysis; select report format

Click OK to confirm.

📋 From manual


Step 3: Start Analysis

With Auto-Sampler

  1. Click Start icon (green play button)
  2. Auto-sampler injects and acquisition begins automatically

Manual Injection

  1. Click Start in software first
  2. Inject sample with syringe (see Manual Injection for technique)
  3. Immediately press the physical START button on the GC

⚠️ Warning: Pressing START on the GC without clicking Start in software first will run the temperature program but no data will be acquired.

📋 From manual


Step 4: Monitor and Complete

During analysis:

  • GC status shows “Acquire”
  • Analysis stops automatically after the preset analysis time
  • To stop early: click Stop icon — data acquired up to that point is saved

After analysis:

  • Status briefly shows “Not Ready”
  • Returns to “Ready” → next sample can be started

📋 From manual